CALIBRATION
NOTE:The TI680 has only one input jack.
Calibrate each heater separately.
For heater A:
1. Using option 1, turn on power and set left
Temperature Control Knob to 400∞F
(205∞C). Allow heater A to warm up.
2. Adjust LO-Temp. Calibration Pot on the
left side of the station so the TI680 Meter
will read 500∞F (260∞C).
If you are using option 2, adjust pot so
meter will read 640∞F (338∞C)
3. Set Temperature Control Knob to 800∞F
(427∞C).
4. Adjust HI-Temp. Calibration Pot so the
Meter will read 900∞F (482∞C).
If you are using option 2, adjust pot so
meter will read 1120∞F (604∞C)
For heater B:
1. Using option 1, set right Temperature
Control Knob to 400∞F (205∞C). Allow
heater B to warm up.
2. Adjust LO-Temp. Calibration Pot on the
right side of the station so the TI680 Meter
will read 500∞F (260∞C).
If you are using option 2, adjust pot so
meter will read 490∞F (254∞C)
3. Set Temperature Control Knob to 800∞F
(427∞C).
4. Adjust HI-Temp. Calibration Pot so the
Meter will read 900∞F (482∞C).
If you are using option 2, adjust pot so
meter will read 980∞F (527∞C)
SOLDERING TIPS
TSX715 .09 .55
Focus Nozzle (2.3 mm) (14.0 mm)
TSX713 .05 .43
Small (1.3 mm) (10.9 mm)
TSX716 .02 .80
Round Needle (.5 mm) (20.3 mm)
TSX717 .01 .63
Small Oval Needle (.3 mm) (16.0 mm)
TSX718 .02 .80
Medium Oval Needle (.5 mm) (20.3 mm)
Hole Dia. Length
TC271
for heater A
TC383
for heater B
heater A
heater B
TI680 meter
option1
option 2
THERMOCOUPLE
PROBES
LO HI
Remove these.
Then install
thermocouple
probes.
TP476
for heater B
TP475
for heater A
cable sold
separately
IMPORTANT:
User must understand the prin-
ciple of wire probe (option 2). See
Form.800. Procedure may yield dif-
ferent temperatures than noted
above.